Long-term stability of the Hounsfield unit to electron density calibration curve in cone-beam computed tomography images for adaptive radiotherapy treatment planning. Issue 4 (15th July 2015)
- Record Type:
- Journal Article
- Title:
- Long-term stability of the Hounsfield unit to electron density calibration curve in cone-beam computed tomography images for adaptive radiotherapy treatment planning. Issue 4 (15th July 2015)
- Main Title:
- Long-term stability of the Hounsfield unit to electron density calibration curve in cone-beam computed tomography images for adaptive radiotherapy treatment planning
- Authors:
- Takemura, Akihiro
Tanabe, Shogo
Tokai, Mei
Ueda, Shinichi
Noto, Kimiya
Isomura, Naoki
Kojima, Hironori - Abstract:
- Abstract: Aim: To use cone-beam computed tomography (CBCT) images for treatment planning, the Hounsfield unit (HU)-electron density (ED) calibration table for CBCT should be stable. The purpose of this study was to verify the stability of the HU values for the CBCT system over 1 year and to evaluate the effects of variation in HU-ED calibration curves on dose calculation. Materials and Methods: A tissue characterisation phantom was scanned with the field of view (FOV) of size S (FOV-S) and FOV of size M (FOV-M) using the CBCT system once a month for 1 year. A single field treatment plan was constructed on digital phantom images to validate the dose distribution using mean HU-ED calibration curves and possible variations. Results: HU values for each material rod over the observation period varied with trend. The HU value of the cortical bone rod decreased by about 100 HU for the FOV-S and by about 300 HU for the FOV-M. Possible variation in the HU-ED calibration curves produced a ≤17·9% dose difference in the dose maximum in the treatment plan. Conclusions: The CBCT system should be calibrated periodically for consistent dose calculation.
- Is Part Of:
- Journal of radiotherapy in practice. Volume 14:Issue 4(2014)
- Journal:
- Journal of radiotherapy in practice
- Issue:
- Volume 14:Issue 4(2014)
- Issue Display:
- Volume 14, Issue 4 (2014)
- Year:
- 2014
- Volume:
- 14
- Issue:
- 4
- Issue Sort Value:
- 2014-0014-0004-0000
- Page Start:
- 410
- Page End:
- 417
- Publication Date:
- 2015-07-15
- Subjects:
- adaptive radiotherapy, -- CBCT, -- dose calculation, -- HU-ED calibration table, -- long-term stability
Radiotherapy -- Periodicals
615.842005 - Journal URLs:
- http://journals.cambridge.org/action/displayJournal?jid=JRP ↗
- DOI:
- 10.1017/S1460396915000321 ↗
- Languages:
- English
- ISSNs:
- 1460-3969
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library HMNTS - ELD Digital store
- Ingest File:
- 2636.xml