Agreement between two oscillometric blood pressure technologies and invasively measured arterial pressure in the dog. (9th October 2015)
- Record Type:
- Journal Article
- Title:
- Agreement between two oscillometric blood pressure technologies and invasively measured arterial pressure in the dog. (9th October 2015)
- Main Title:
- Agreement between two oscillometric blood pressure technologies and invasively measured arterial pressure in the dog
- Authors:
- da Cunha, Anderson F
Ramos, Sara J
Domingues, Michelle
Beaufrère, Hugues
Shelby, Amanda
Stout, Rhett
Acierno, Mark J - Abstract:
- Abstract: Objectives: To compare two commonly used oscillometric technologies for obtaining noninvasive blood pressure (NIBP) measurements and to determine if there is a difference in agreement between these systems and invasive blood pressure (IBP) measurements. Study design: Prospective, experimental study. Animals: Twenty adult laboratory dogs. Methods: Each dog was anesthetized and its median caudal artery catheterized for IBP monitoring. An NIBP cuff was placed in the middle third of the antebrachium and attached to either monitor‐1 or monitor‐2. Four pairs of concurrent NIBP and IBP measurements were recorded with each monitor. Agreement between IBP and NIBP measurements was explored using Bland–Altman analysis, as well as the American College of Veterinary Internal Medicine (ACVIM) and Association for the Advancement of Medical Instrumentation (AAMI) guidelines for the validation of NIBP devices. Results: Both NIBP technologies produced results that met the ACVIM and AAMI guidelines for the validation of NIBP devices. For monitor‐1, analyses of agreement showed biases of 0.2 mmHg [95% limits of agreement (LoA) −11.8 to 12.3 mmHg] in systolic arterial pressure (SAP) values, −2.6 mmHg (95% LoA −14.4 to 9.1 mmHg) in diastolic arterial pressure (DAP) values, and −2.5 mmHg (95% LoA −12.7 to 7.3 mmHg) in mean arterial pressure (MAP) values. For monitor‐2, analyses of agreement showed biases of 3.4 mmHg (95% LoA −8.7 to 15.5 mmHg) in SAP values, 2.2 mmHg (95% LoA −6.6 toAbstract: Objectives: To compare two commonly used oscillometric technologies for obtaining noninvasive blood pressure (NIBP) measurements and to determine if there is a difference in agreement between these systems and invasive blood pressure (IBP) measurements. Study design: Prospective, experimental study. Animals: Twenty adult laboratory dogs. Methods: Each dog was anesthetized and its median caudal artery catheterized for IBP monitoring. An NIBP cuff was placed in the middle third of the antebrachium and attached to either monitor‐1 or monitor‐2. Four pairs of concurrent NIBP and IBP measurements were recorded with each monitor. Agreement between IBP and NIBP measurements was explored using Bland–Altman analysis, as well as the American College of Veterinary Internal Medicine (ACVIM) and Association for the Advancement of Medical Instrumentation (AAMI) guidelines for the validation of NIBP devices. Results: Both NIBP technologies produced results that met the ACVIM and AAMI guidelines for the validation of NIBP devices. For monitor‐1, analyses of agreement showed biases of 0.2 mmHg [95% limits of agreement (LoA) −11.8 to 12.3 mmHg] in systolic arterial pressure (SAP) values, −2.6 mmHg (95% LoA −14.4 to 9.1 mmHg) in diastolic arterial pressure (DAP) values, and −2.5 mmHg (95% LoA −12.7 to 7.3 mmHg) in mean arterial pressure (MAP) values. For monitor‐2, analyses of agreement showed biases of 3.4 mmHg (95% LoA −8.7 to 15.5 mmHg) in SAP values, 2.2 mmHg (95% LoA −6.6 to 10.9 mmHg) in DAP values, and 1.6 mmHg (95% LoA −5.9 to 8.9 mmHg) in MAP values. Conclusions and clinical relevance: Multi‐function monitors can contain components from various manufacturers. Clinicians should consider whether these have been validated in the species to be monitored. Both of the technologies studied here seem appropriate for use in dogs. … (more)
- Is Part Of:
- Veterinary anaesthesia and analgesia. Volume 43:Number 2(2016:Mar.)
- Journal:
- Veterinary anaesthesia and analgesia
- Issue:
- Volume 43:Number 2(2016:Mar.)
- Issue Display:
- Volume 43, Issue 2 (2016)
- Year:
- 2016
- Volume:
- 43
- Issue:
- 2
- Issue Sort Value:
- 2016-0043-0002-0000
- Page Start:
- 199
- Page End:
- 203
- Publication Date:
- 2015-10-09
- Subjects:
- canine -- comparison -- IBP -- monitoring -- NIBP
Veterinary anesthesia -- Periodicals
636.089 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1467-2995 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1111/vaa.12312 ↗
- Languages:
- English
- ISSNs:
- 1467-2987
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 9226.528500
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 2336.xml