Longevity of the Subcutaneous Implantable Defibrillator: Long-Term Follow-Up of the European Regulatory Trial Cohort. (October 2015)
- Record Type:
- Journal Article
- Title:
- Longevity of the Subcutaneous Implantable Defibrillator: Long-Term Follow-Up of the European Regulatory Trial Cohort. (October 2015)
- Main Title:
- Longevity of the Subcutaneous Implantable Defibrillator
- Authors:
- Theuns, Dominic A.M.J.
Crozier, Ian G.
Barr, Craig S.
Hood, Margaret A.
Cappato, Riccardo
Knops, Reinoud E.
Maass, Alexander H.
Boersma, Lucas V.A.
Jordaens, Luc - Abstract:
- Abstract : Background—: The recent advent of subcutaneous implantable cardioverter defibrillators (S-ICDs) has provided investigators with a safe and effective new therapy in patients at risk of sudden cardiac death. At present, no data are available with regard to the longevity of these new devices. This study evaluated the longevity of the S-ICD system. Methods and Results—: All patients enrolled in the European Regulatory Trial were included in the analysis. During follow-up, time and causes of device replacement or explantation were assessed and categorized. Device longevity was estimated using Kaplan–Meier analysis. Fifty-five patients were followed for a median of 5.8 years. During follow-up, 26 (47%) patients underwent device replacement and 5 (9%) underwent device explantation. Median time to replacement was 5.0 years (Q1–Q3, 4.4–5.6 years). Replacement was caused by battery depletion in 25 patients (92%), of which 5 within 1.5 years because of premature battery depletion, and by infection in 1 patient (2%). Replacement for a transvenous ICD system was required in 4 patients (7%) because of ineffective defibrillation in 1 (0.003 per patient-year), need for resynchronization therapy in 2 (0.01 per patient-year), and for antibradycardia pacing in 1 (0.003 per patient-year). At 5 years follow-up, 71% of devices were still in service. Conclusions—: This study provides the first estimate of S-ICD system longevity since its introduction in clinical practice. MedianAbstract : Background—: The recent advent of subcutaneous implantable cardioverter defibrillators (S-ICDs) has provided investigators with a safe and effective new therapy in patients at risk of sudden cardiac death. At present, no data are available with regard to the longevity of these new devices. This study evaluated the longevity of the S-ICD system. Methods and Results—: All patients enrolled in the European Regulatory Trial were included in the analysis. During follow-up, time and causes of device replacement or explantation were assessed and categorized. Device longevity was estimated using Kaplan–Meier analysis. Fifty-five patients were followed for a median of 5.8 years. During follow-up, 26 (47%) patients underwent device replacement and 5 (9%) underwent device explantation. Median time to replacement was 5.0 years (Q1–Q3, 4.4–5.6 years). Replacement was caused by battery depletion in 25 patients (92%), of which 5 within 1.5 years because of premature battery depletion, and by infection in 1 patient (2%). Replacement for a transvenous ICD system was required in 4 patients (7%) because of ineffective defibrillation in 1 (0.003 per patient-year), need for resynchronization therapy in 2 (0.01 per patient-year), and for antibradycardia pacing in 1 (0.003 per patient-year). At 5 years follow-up, 71% of devices were still in service. Conclusions—: This study provides the first estimate of S-ICD system longevity since its introduction in clinical practice. Median longevity of the first generation S-ICD system was 5.0 years. The majority of devices were replaced because of battery depletion. Clinical Trial Registration—: URL:http://www.clinicaltrials.gov . Unique identifier: NCT01117792. … (more)
- Is Part Of:
- Circulation. Volume 8:Number 5(2015)
- Journal:
- Circulation
- Issue:
- Volume 8:Number 5(2015)
- Issue Display:
- Volume 8, Issue 5 (2015)
- Year:
- 2015
- Volume:
- 8
- Issue:
- 5
- Issue Sort Value:
- 2015-0008-0005-0000
- Page Start:
- Page End:
- Publication Date:
- 2015-10
- Subjects:
- arrhythmias, cardiac -- bradycardia -- death, sudden, cardiac -- defibrillators, implantable -- longevity
Arrhythmia -- Periodicals
Heart -- Electric properties -- Periodicals
616.128 - Journal URLs:
- http://gateway.ovid.com/ovidweb.cgi?T=JS&MODE=ovid&NEWS=n&PAGE=toc&D=ovft&AN=01337493-000000000-00000 ↗
http://circep.ahajournals.org/ ↗
http://journals.lww.com ↗ - DOI:
- 10.1161/CIRCEP.115.002953 ↗
- Languages:
- English
- ISSNs:
- 1941-3149
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3265.262500
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File: