Influence of temporal regularization and radial undersampling factor on compressed sensing reconstruction in dynamic contrast enhanced MRI of the breast. Issue 1 (1st June 2015)
- Record Type:
- Journal Article
- Title:
- Influence of temporal regularization and radial undersampling factor on compressed sensing reconstruction in dynamic contrast enhanced MRI of the breast. Issue 1 (1st June 2015)
- Main Title:
- Influence of temporal regularization and radial undersampling factor on compressed sensing reconstruction in dynamic contrast enhanced MRI of the breast
- Authors:
- Kim, Sungheon G.
Feng, Li
Grimm, Robert
Freed, Melanie
Block, Kai Tobias
Sodickson, Daniel K.
Moy, Linda
Otazo, Ricardo - Abstract:
- Abstract : Background: To evaluate the influence of temporal sparsity regularization and radial undersampling on compressed sensing reconstruction of dynamic contrast‐enhanced (DCE) MRI, using the iterative Golden‐angle RAdial Sparse Parallel (iGRASP) MRI technique in the setting of breast cancer evaluation. Methods: DCE‐MRI examinations of the breast (n = 7) were conducted using iGRASP at 3 Tesla. Images were reconstructed with five different radial undersampling schemes corresponding to temporal resolutions between 2 and 13.4 s/frame and with four different weights for temporal sparsity regularization (λ = 0.1, 0.5, 2, and 6 times of noise level). Image similarity to time‐averaged reference images was assessed by two breast radiologists and using quantitative metrics. Temporal similarity was measured in terms of wash‐in slope and contrast kinetic model parameters. Results: iGRASP images reconstructed with λ = 2 and 5.1 s/frame had significantly ( P < 0.05) higher similarity to time‐averaged reference images than the images with other reconstruction parameters (mutual information (MI) >5%), in agreement with the assessment of two breast radiologists. Higher undersampling (temporal resolution < 5.1 s/frame) required stronger temporal sparsity regularization (λ ≥ 2) to remove streaking aliasing artifacts (MI > 23% between λ = 2 and 0.5). The difference between the kinetic‐model transfer rates of benign and malignant groups decreased as temporal resolution decreased (82%Abstract : Background: To evaluate the influence of temporal sparsity regularization and radial undersampling on compressed sensing reconstruction of dynamic contrast‐enhanced (DCE) MRI, using the iterative Golden‐angle RAdial Sparse Parallel (iGRASP) MRI technique in the setting of breast cancer evaluation. Methods: DCE‐MRI examinations of the breast (n = 7) were conducted using iGRASP at 3 Tesla. Images were reconstructed with five different radial undersampling schemes corresponding to temporal resolutions between 2 and 13.4 s/frame and with four different weights for temporal sparsity regularization (λ = 0.1, 0.5, 2, and 6 times of noise level). Image similarity to time‐averaged reference images was assessed by two breast radiologists and using quantitative metrics. Temporal similarity was measured in terms of wash‐in slope and contrast kinetic model parameters. Results: iGRASP images reconstructed with λ = 2 and 5.1 s/frame had significantly ( P < 0.05) higher similarity to time‐averaged reference images than the images with other reconstruction parameters (mutual information (MI) >5%), in agreement with the assessment of two breast radiologists. Higher undersampling (temporal resolution < 5.1 s/frame) required stronger temporal sparsity regularization (λ ≥ 2) to remove streaking aliasing artifacts (MI > 23% between λ = 2 and 0.5). The difference between the kinetic‐model transfer rates of benign and malignant groups decreased as temporal resolution decreased (82% between 2 and 13.4 s/frame). Conclusion: This study demonstrates objective spatial and temporal similarity measures can be used to assess the influence of sparsity constraint and undersampling in compressed sensing DCE‐MRI and also shows that the iGRASP method provides the flexibility of optimizing these reconstruction parameters in the postprocessing stage using the same acquired data. J. MAGN. RESON. IMAGING 2016;43:261–269. … (more)
- Is Part Of:
- Journal of magnetic resonance imaging. Volume 43:Issue 1(2016)
- Journal:
- Journal of magnetic resonance imaging
- Issue:
- Volume 43:Issue 1(2016)
- Issue Display:
- Volume 43, Issue 1 (2016)
- Year:
- 2016
- Volume:
- 43
- Issue:
- 1
- Issue Sort Value:
- 2016-0043-0001-0000
- Page Start:
- 261
- Page End:
- 269
- Publication Date:
- 2015-06-01
- Subjects:
- compressed sensing -- parallel imaging -- temporal sparsity -- radial undersampling -- golden‐angle -- DCE‐MRI -- breast cancer -- iGRASP
Magnetic resonance imaging -- Periodicals
616 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1522-2586 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/jmri.24961 ↗
- Languages:
- English
- ISSNs:
- 1053-1807
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5010.791000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 893.xml