Mindray BC‐6800 body fluid mode, performance of nucleated cells, and differential count in ascitic and pleural fluids. (7th November 2015)
- Record Type:
- Journal Article
- Title:
- Mindray BC‐6800 body fluid mode, performance of nucleated cells, and differential count in ascitic and pleural fluids. (7th November 2015)
- Main Title:
- Mindray BC‐6800 body fluid mode, performance of nucleated cells, and differential count in ascitic and pleural fluids
- Authors:
- Buoro, S.
Mecca, T.
Azzarà, G.
Seghezzi, M.
Candiago, E.
Gianatti, A.
Crippa, A.
La Gioia, A. - Abstract:
- Summary: Introduction: An accurate and rapid analysis of cells in body fluids (BFs) is important for diagnosis and follow‐up in many pathological conditions. We evaluated the analytical performance of the module BF Mindray BC‐6800 (BC‐6800‐BF) for cytometric analysis of ascitic and pleural fluids. Methods: A total of 99 ascitic and 45 pleural samples were collected and assessed with BC‐6800‐BF and optical microscopy. This study also includes the evaluation of limit blank (LoB), limit detection (LoD), limit quantitation, (LoQ), carryover, linearity, and diagnostic concordance between the two methods. Results: For TC‐BF, LoB was 1 × 10 6 cells/L, LoD was 3 × 10 6 cells/L, and LoQ was 4 × 10 6 cells/L. Linearity was excellent ( r 2 = 0.99) and carryover was negligible. TC‐BF performed with the two methods showed Pearson's correlation of 0.99 ( P < 0.0001), Passing–Bablok regression y = 1.04 x − 1.17, and bias 33.7 cells. In ascitic fluids, polymorphonuclear cells (PMN) showed an area under curve (AUC) of 0.98 ( P < 0.0001). In pleural fluids, mononuclear cells (MN) and PMN % displayed an AUC of 0.79 ( P < 0.0001) and 0.93 ( P < 0.0001), respectively. Conclusions: BC‐6800‐BF in ascitic and pleural fluids offers rapid and accurate cell and differential counts in clinically relevant concentration ranges. The use of BC‐6800‐BF may allow to replace routine optical counting, except for samples displaying abnormal cell counts or abnormal DIFF scattergram.
- Is Part Of:
- International journal of laboratory hematology. Volume 38:Number 1(2016:Feb.)
- Journal:
- International journal of laboratory hematology
- Issue:
- Volume 38:Number 1(2016:Feb.)
- Issue Display:
- Volume 38, Issue 1 (2016)
- Year:
- 2016
- Volume:
- 38
- Issue:
- 1
- Issue Sort Value:
- 2016-0038-0001-0000
- Page Start:
- 90
- Page End:
- 101
- Publication Date:
- 2015-11-07
- Subjects:
- Ascitic fluid -- pleural fluid -- automated cell counting -- body fluid -- Mindray BC‐6800
Hematology -- Periodicals
Blood -- Diseases -- Periodicals
Hematology -- Periodicals
616.15005 - Journal URLs:
- http://firstsearch.oclc.org/FSIP?db=ECO&journal=1751-5521&screen=info&done=referer ↗
http://www.blackwell-synergy.com/loi/clh ↗
http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1751-553X ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1111/ijlh.12446 ↗
- Languages:
- English
- ISSNs:
- 1751-5521
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4542.312220
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 378.xml