A multicentre 'end to end' dosimetry audit for cervix HDR brachytherapy treatment. Issue 2 (February 2015)
- Record Type:
- Journal Article
- Title:
- A multicentre 'end to end' dosimetry audit for cervix HDR brachytherapy treatment. Issue 2 (February 2015)
- Main Title:
- A multicentre 'end to end' dosimetry audit for cervix HDR brachytherapy treatment
- Authors:
- Palmer, Antony L.
Diez, Patricia
Gandon, Laura
Wynn-Jones, Andrea
Bownes, Peter
Lee, Chris
Aird, Edwin
Bidmead, Margaret
Lowe, Gerry
Bradley, David
Nisbet, Andrew - Abstract:
- <abstract xml:lang="en" abstract-type="author" id="ab005"> <title id="st060">Abstract</title> <sec> <title id="st065">Purpose</title> <p id="sp0005">To undertake the first multicentre fully 'end to end' dosimetry audit for HDR cervix brachytherapy, comparing planned and delivered dose distributions around clinical treatment applicators, with review of local procedures.</p> </sec> <sec> <title id="st070">Materials and methods</title> <p id="sp0010">A film-dosimetry audit was performed at 46 centres, including imaging, applicator reconstruction, treatment planning and delivery. Film dose maps were calculated using triple-channel dosimetry and compared to RTDose data from treatment planning systems. Deviations between plan and measurement were quantified at prescription Point A and using gamma analysis. Local procedures were also discussed.</p> </sec> <sec> <title id="st075">Results</title> <p id="sp0015">The mean difference between planned and measured dose at Point A was −0.6% for plastic applicators and −3.0% for metal applicators, at standard uncertainty 3.0% (<italic>k</italic> = 1). Isodose distributions agreed within 1 mm over a dose range 2–16 Gy. Mean gamma passing rates exceeded 97% for plastic and metal applicators at 3% (local) 2 mm criteria. Two errors were found: one dose normalisation error and one applicator library misaligned with the imaged applicator. Suggestions for quality improvement were also made.</p> </sec> <sec> <title id="st080">Conclusions</title> <p<abstract xml:lang="en" abstract-type="author" id="ab005"> <title id="st060">Abstract</title> <sec> <title id="st065">Purpose</title> <p id="sp0005">To undertake the first multicentre fully 'end to end' dosimetry audit for HDR cervix brachytherapy, comparing planned and delivered dose distributions around clinical treatment applicators, with review of local procedures.</p> </sec> <sec> <title id="st070">Materials and methods</title> <p id="sp0010">A film-dosimetry audit was performed at 46 centres, including imaging, applicator reconstruction, treatment planning and delivery. Film dose maps were calculated using triple-channel dosimetry and compared to RTDose data from treatment planning systems. Deviations between plan and measurement were quantified at prescription Point A and using gamma analysis. Local procedures were also discussed.</p> </sec> <sec> <title id="st075">Results</title> <p id="sp0015">The mean difference between planned and measured dose at Point A was −0.6% for plastic applicators and −3.0% for metal applicators, at standard uncertainty 3.0% (<italic>k</italic> = 1). Isodose distributions agreed within 1 mm over a dose range 2–16 Gy. Mean gamma passing rates exceeded 97% for plastic and metal applicators at 3% (local) 2 mm criteria. Two errors were found: one dose normalisation error and one applicator library misaligned with the imaged applicator. Suggestions for quality improvement were also made.</p> </sec> <sec> <title id="st080">Conclusions</title> <p id="sp0020">The concept of 'end to end' dosimetry audit for HDR brachytherapy has been successfully implemented in a multicentre environment, providing evidence that a high level of accuracy in brachytherapy dosimetry can be achieved.</p> </sec> </abstract> … (more)
- Is Part Of:
- Radiotherapy and oncology. Volume 114:Issue 2(2015:Feb.)
- Journal:
- Radiotherapy and oncology
- Issue:
- Volume 114:Issue 2(2015:Feb.)
- Issue Display:
- Volume 114, Issue 2 (2015)
- Year:
- 2015
- Volume:
- 114
- Issue:
- 2
- Issue Sort Value:
- 2015-0114-0002-0000
- Page Start:
- 264
- Page End:
- 271
- Publication Date:
- 2015-02
- Subjects:
- Oncology -- Periodicals
Radiotherapy -- Periodicals
Tumors -- Periodicals
Medical Oncology -- Periodicals
Neoplasms -- radiotherapy -- Periodicals
Radiotherapy -- Periodicals
Radiothérapie -- Périodiques
Cancérologie -- Périodiques
Tumeurs -- Périodiques
Electronic journals
616.9940642 - Journal URLs:
- http://www.sciencedirect.com/science/journal/01678140 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/01678140 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/01678140 ↗
http://www.estro.org/ ↗
http://www.elsevier.com/journals ↗
http://www.journals.elsevier.com/radiotherapy-and-oncology/ ↗ - DOI:
- 10.1016/j.radonc.2014.12.006 ↗
- Languages:
- English
- ISSNs:
- 0167-8140
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 7240.790000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 4111.xml