In vitro biomechanical evaluation of four fixation techniques for distractive–flexion injury stage 3 of the cervical spine. (August 2015)
- Record Type:
- Journal Article
- Title:
- In vitro biomechanical evaluation of four fixation techniques for distractive–flexion injury stage 3 of the cervical spine. (August 2015)
- Main Title:
- In vitro biomechanical evaluation of four fixation techniques for distractive–flexion injury stage 3 of the cervical spine
- Authors:
- Henriques, Thomas
Cunningham, Bryan W.
Mcafee, Paul C.
Olerud, Claes - Abstract:
- <abstract> <title>Abstract</title> <p> <bold> <italic>Purpose.</italic> </bold> Anterior plate fixation has been reported to provide satisfactory results in cervical spine distractive flexion (DF) injuries stages 1 and 2, but will result in a substantial failure rate in more unstable stage 3 and above. The aim of this investigation was to determine the biomechanical properties of different fixation techniques in a DF-3 injury model where all structures responsible for the posterior tension band mechanism are torn.</p> <p> <italic>Methods.</italic> The multidirectional three-dimensional stiffness of the subaxial cervical spine was measured in eight cadaveric specimens with a simulated DF-3 injury at C5–C6, stabilized with four different fixation techniques: anterior plate alone, anterior plate combined with posterior wire, transarticular facet screws, and a pedicle screw–rod construct, respectively.</p> <p> <italic>Results.</italic> The anterior plate alone did not improve stability compared to the intact spine condition, thus allowing considerable range of motion around all three cardinal axes (<italic>p</italic> &gt; 0.05). The anterior plate combined with posterior wire technique improved flexion–extension stiffness (<italic>p</italic> = 0.023), but not in axial rotation and lateral bending. When the anterior plate was combined with transarticular facet screws or with a pedicle screws–rod instrumentation, the stability improved in flexion–extension, lateral bending, and in<abstract> <title>Abstract</title> <p> <bold> <italic>Purpose.</italic> </bold> Anterior plate fixation has been reported to provide satisfactory results in cervical spine distractive flexion (DF) injuries stages 1 and 2, but will result in a substantial failure rate in more unstable stage 3 and above. The aim of this investigation was to determine the biomechanical properties of different fixation techniques in a DF-3 injury model where all structures responsible for the posterior tension band mechanism are torn.</p> <p> <italic>Methods.</italic> The multidirectional three-dimensional stiffness of the subaxial cervical spine was measured in eight cadaveric specimens with a simulated DF-3 injury at C5–C6, stabilized with four different fixation techniques: anterior plate alone, anterior plate combined with posterior wire, transarticular facet screws, and a pedicle screw–rod construct, respectively.</p> <p> <italic>Results.</italic> The anterior plate alone did not improve stability compared to the intact spine condition, thus allowing considerable range of motion around all three cardinal axes (<italic>p</italic> &gt; 0.05). The anterior plate combined with posterior wire technique improved flexion–extension stiffness (<italic>p</italic> = 0.023), but not in axial rotation and lateral bending. When the anterior plate was combined with transarticular facet screws or with a pedicle screws–rod instrumentation, the stability improved in flexion–extension, lateral bending, and in axial rotation (<italic>p</italic> &lt; 0.05).</p> <p> <italic>Conclusions.</italic> These findings imply that the use of anterior fixation alone is insufficient for fixation of the highly unstable DF-3 injury. In these situations, the use of anterior fixation combined with a competent posterior tension band reconstruction (e.g. transarticular screws or a posterior pedicle screws–rod device) improves segmental stability.</p> </abstract> … (more)
- Is Part Of:
- Upsala journal of medical sciences. Volume 120:Number 3(2015:Sep.)
- Journal:
- Upsala journal of medical sciences
- Issue:
- Volume 120:Number 3(2015:Sep.)
- Issue Display:
- Volume 120, Issue 3 (2015)
- Year:
- 2015
- Volume:
- 120
- Issue:
- 3
- Issue Sort Value:
- 2015-0120-0003-0000
- Page Start:
- 198
- Page End:
- 206
- Publication Date:
- 2015-08
- Subjects:
- Medicine -- Periodicals
Medicine -- Periodicals
Electronic journals
610.5 - Journal URLs:
- http://informahealthcare.com/loi/ups ↗
http://informahealthcare.com/ups ↗
http://search.epnet.com/direct.asp?db=aph&jn="BE3"&scope=site ↗
https://ujms.net/index.php/ujms/issue/view/523 ↗
http://informahealthcare.com ↗
http://www.tandf.co.uk/journals/titles/03009734.html ↗ - DOI:
- ↗
- Languages:
- English
- ISSNs:
- 0300-9734
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 9122.700000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 3655.xml