Accuracy of applicator tip reconstruction in MRI-guided interstitial 192Ir-high-dose-rate brachytherapy of liver tumors. Issue 1 (April 2015)
- Record Type:
- Journal Article
- Title:
- Accuracy of applicator tip reconstruction in MRI-guided interstitial 192Ir-high-dose-rate brachytherapy of liver tumors. Issue 1 (April 2015)
- Main Title:
- Accuracy of applicator tip reconstruction in MRI-guided interstitial 192Ir-high-dose-rate brachytherapy of liver tumors
- Authors:
- Wybranski, Christian
Eberhardt, Benjamin
Fischbach, Katharina
Fischbach, Frank
Walke, Mathias
Hass, Peter
Röhl, Friedrich-Wilhelm
Kosiek, Ortrud
Kaiser, Mandy
Pech, Maciej
Lüdemann, Lutz
Ricke, Jens - Abstract:
- <abstract xml:lang="en" abstract-type="author" id="ab005"> <title id="st120">Abstract</title> <sec> <title id="st080">Background and purpose</title> <p id="sp0005">To evaluate the reconstruction accuracy of brachytherapy (BT) applicators tips in vitro and in vivo in MRI-guided <sup>192</sup>Ir-high-dose-rate (HDR)-BT of inoperable liver tumors.</p> </sec> <sec> <title id="st085">Materials and methods</title> <p id="sp0010">Reconstruction accuracy of plastic BT applicators, visualized by nitinol inserts, was assessed in MRI phantom measurements and in MRI <sup>192</sup>Ir-HDR-BT treatment planning datasets of 45 patients employing CT co-registration and vector decomposition. Conspicuity, short-term dislocation, and reconstruction errors were assessed in the clinical data. The clinical effect of applicator reconstruction accuracy was determined in follow-up MRI data.</p> </sec> <sec> <title id="st090">Results</title> <p id="sp0015">Applicator reconstruction accuracy was 1.6 ± 0.5 mm in the phantom measurements. In the clinical MRI datasets applicator conspicuity was rated good/optimal in ⩾72% of cases. 16/129 applicators showed not time dependent deviation in between MRI/CT acquisition (<italic>p</italic> &gt; 0.1). Reconstruction accuracy was 5.5 ± 2.8 mm, and the average image co-registration error was 3.1 ± 0.9 mm. Vector decomposition revealed no preferred direction of reconstruction errors. In the follow-up data deviation of planned dose distribution and irradiation<abstract xml:lang="en" abstract-type="author" id="ab005"> <title id="st120">Abstract</title> <sec> <title id="st080">Background and purpose</title> <p id="sp0005">To evaluate the reconstruction accuracy of brachytherapy (BT) applicators tips in vitro and in vivo in MRI-guided <sup>192</sup>Ir-high-dose-rate (HDR)-BT of inoperable liver tumors.</p> </sec> <sec> <title id="st085">Materials and methods</title> <p id="sp0010">Reconstruction accuracy of plastic BT applicators, visualized by nitinol inserts, was assessed in MRI phantom measurements and in MRI <sup>192</sup>Ir-HDR-BT treatment planning datasets of 45 patients employing CT co-registration and vector decomposition. Conspicuity, short-term dislocation, and reconstruction errors were assessed in the clinical data. The clinical effect of applicator reconstruction accuracy was determined in follow-up MRI data.</p> </sec> <sec> <title id="st090">Results</title> <p id="sp0015">Applicator reconstruction accuracy was 1.6 ± 0.5 mm in the phantom measurements. In the clinical MRI datasets applicator conspicuity was rated good/optimal in ⩾72% of cases. 16/129 applicators showed not time dependent deviation in between MRI/CT acquisition (<italic>p</italic> &gt; 0.1). Reconstruction accuracy was 5.5 ± 2.8 mm, and the average image co-registration error was 3.1 ± 0.9 mm. Vector decomposition revealed no preferred direction of reconstruction errors. In the follow-up data deviation of planned dose distribution and irradiation effect was 6.9 ± 3.3 mm matching the mean co-registration error (6.5 ± 2.5 mm; <italic>p</italic> &gt; 0.1).</p> </sec> <sec> <title id="st095">Conclusion</title> <p id="sp0020">Applicator reconstruction accuracy in vitro conforms to AAPM TG 56 standard. Nitinol-inserts are feasible for applicator visualization and yield good conspicuity in MRI treatment planning data. No preferred direction of reconstruction errors were found in vivo.</p> </sec> </abstract> … (more)
- Is Part Of:
- Radiotherapy and oncology. Volume 115:Issue 1(2015:Apr.)
- Journal:
- Radiotherapy and oncology
- Issue:
- Volume 115:Issue 1(2015:Apr.)
- Issue Display:
- Volume 115, Issue 1 (2015)
- Year:
- 2015
- Volume:
- 115
- Issue:
- 1
- Issue Sort Value:
- 2015-0115-0001-0000
- Page Start:
- 72
- Page End:
- 77
- Publication Date:
- 2015-04
- Subjects:
- Oncology -- Periodicals
Radiotherapy -- Periodicals
Tumors -- Periodicals
Medical Oncology -- Periodicals
Neoplasms -- radiotherapy -- Periodicals
Radiotherapy -- Periodicals
Radiothérapie -- Périodiques
Cancérologie -- Périodiques
Tumeurs -- Périodiques
Electronic journals
616.9940642 - Journal URLs:
- http://www.sciencedirect.com/science/journal/01678140 ↗
http://www.clinicalkey.com/dura/browse/journalIssue/01678140 ↗
http://www.clinicalkey.com.au/dura/browse/journalIssue/01678140 ↗
http://www.estro.org/ ↗
http://www.elsevier.com/journals ↗
http://www.journals.elsevier.com/radiotherapy-and-oncology/ ↗ - DOI:
- 10.1016/j.radonc.2015.01.018 ↗
- Languages:
- English
- ISSNs:
- 0167-8140
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 7240.790000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 3969.xml