Interference of GSM mobile phones with communication between Cardiac Rhythm Management devices and programmers: A combined in vivo and in vitro study. Issue 5 (10th April 2015)
- Record Type:
- Journal Article
- Title:
- Interference of GSM mobile phones with communication between Cardiac Rhythm Management devices and programmers: A combined in vivo and in vitro study. Issue 5 (10th April 2015)
- Main Title:
- Interference of GSM mobile phones with communication between Cardiac Rhythm Management devices and programmers: A combined in vivo and in vitro study
- Authors:
- Huang, Dong
Dong, Zhi‐Feng
Chen, Yan
Wang, Fa‐Bin
Wei, Zhi
Zhao, Wen‐Bin
Li, Shuai
Liu, Ming‐Ya
Zhu, Wei
Wei, Meng
Li, Jing‐Bo - Abstract:
- <abstract abstract-type="main" xml:lang="en"> <title> <x xml:space="preserve">Abstract</x> </title> <sec id="bem21911-sec-0001" sec-type="section"> <p>To investigate interference, and how to avoid it, by high‐frequency electromagnetic fields (EMFs) of Global System for Mobile Communications (GSM) mobile phone with communication between cardiac rhythm management devices (CRMs) and programmers, a combined in vivo and in vitro testing was conducted. During in vivo testing, GSM mobile phones interfered with CRM‐programmer communication in 33 of 65 subjects tested (50.8%). Losing ventricle sensing was representative in this study. In terms of clinical symptoms, only 4 subjects (0.6%) felt dizzy during testing. CRM‐programmer communication recovered upon termination of mobile phone communication. During in vitro testing, electromagnetic interference by high‐frequency (700–950 MHz) EMFs reproducibly occurred in duplicate testing in 18 of 20 CRMs (90%). During each interference, the pacing pulse signal on the programmer would suddenly disappear while the synchronous signal was normal on the amplifier‐oscilloscope. Simulation analysis showed that interference by radiofrequency emitting devices with CRM‐programmer communication may be attributed to factors including materials, excitation source distance, and implant depth. Results suggested that patients implanted with CRMs should not be restricted from using GSM mobile phones; however, CRMs should be kept away from high‐frequency<abstract abstract-type="main" xml:lang="en"> <title> <x xml:space="preserve">Abstract</x> </title> <sec id="bem21911-sec-0001" sec-type="section"> <p>To investigate interference, and how to avoid it, by high‐frequency electromagnetic fields (EMFs) of Global System for Mobile Communications (GSM) mobile phone with communication between cardiac rhythm management devices (CRMs) and programmers, a combined in vivo and in vitro testing was conducted. During in vivo testing, GSM mobile phones interfered with CRM‐programmer communication in 33 of 65 subjects tested (50.8%). Losing ventricle sensing was representative in this study. In terms of clinical symptoms, only 4 subjects (0.6%) felt dizzy during testing. CRM‐programmer communication recovered upon termination of mobile phone communication. During in vitro testing, electromagnetic interference by high‐frequency (700–950 MHz) EMFs reproducibly occurred in duplicate testing in 18 of 20 CRMs (90%). During each interference, the pacing pulse signal on the programmer would suddenly disappear while the synchronous signal was normal on the amplifier‐oscilloscope. Simulation analysis showed that interference by radiofrequency emitting devices with CRM‐programmer communication may be attributed to factors including materials, excitation source distance, and implant depth. Results suggested that patients implanted with CRMs should not be restricted from using GSM mobile phones; however, CRMs should be kept away from high‐frequency EMFs of GSM mobile phone during programming. Bioelectromagnetics. 36:367–376, 2015. © 2015 Wiley Periodicals, Inc.</p> </sec> </abstract> … (more)
- Is Part Of:
- Bioelectromagnetics. Volume 36:Issue 5(2015:Jul.)
- Journal:
- Bioelectromagnetics
- Issue:
- Volume 36:Issue 5(2015:Jul.)
- Issue Display:
- Volume 36, Issue 5 (2015)
- Year:
- 2015
- Volume:
- 36
- Issue:
- 5
- Issue Sort Value:
- 2015-0036-0005-0000
- Page Start:
- 367
- Page End:
- 376
- Publication Date:
- 2015-04-10
- Subjects:
- Electromagnetism -- Physiological effect -- Periodicals
571.47 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1521-186X ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/bem.21911 ↗
- Languages:
- English
- ISSNs:
- 0197-8462
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 2072.009000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 3510.xml