Feasibility and Efficacy of Percutaneously Delivered Leadless Cardiac Pacing in an In Vivo Ovine Model. (5th December 2014)
- Record Type:
- Journal Article
- Title:
- Feasibility and Efficacy of Percutaneously Delivered Leadless Cardiac Pacing in an In Vivo Ovine Model. (5th December 2014)
- Main Title:
- Feasibility and Efficacy of Percutaneously Delivered Leadless Cardiac Pacing in an In Vivo Ovine Model
- Authors:
- KORUTH, JACOB S.
RIPPY, MARIAN K.
KHAIRKHAHAN, ALEXANDER
LIGON, DAVID A.
HUBBARD, CHRISTOPHER A.
GOAR, FREDERICK ST.
LEE, RANDALL
RIBEIRO, LEONARDO
MILLER, MARC
DUKKIPATI, SRINIVAS
NEUZIL, PETR
REDDY, VIVEK Y. - Abstract:
- <abstract abstract-type="main"> <title>Preclinical Feasibility of a Leadless Pacemaker</title> <sec id="jce12579-sec-0010" sec-type="section"> <title>Introduction</title> <p>In this <italic>in vivo</italic> ovine study, we describe the course of pacing and sensing parameters during follow‐up as well as the gross and histopathological features at the implant site of the recently described leadless cardiac pacemaker (LCP).</p> </sec> <sec id="jce12579-sec-0020" sec-type="section"> <title>Methods and Results</title> <p>All sheep underwent LCP implantation in the right ventricular (RV) apex. Serial pacing/sensing thresholds were assessed. In the first cohort, 11 animals were followed up for a period of 3 months, followed by necropsy. In the second cohort, 7 additional sheep were followed for a period of up to 18 months. Mean pacing and sensing thresholds in the 3‐month cohort were 1.0 ± 0.5 V and 9.0 ± 3.7 mV at implant, and 0.7 ± 0.2 V and 8.1 ± 3.9 mV at 90 days, respectively. At necropsy at 3 months, all devices were securely implanted at the RV apex without extrusion of helix beyond the RV wall. Besides endocardial reactive changes in the RV apex surrounding the distal portion of the LCP, there were no other grossly visible abnormalities. In the second cohort (7 sheep), mean pacing and sensing thresholds were 1.0 ± 0.5 V and 9.9 ± 3.8 mV at implant, and 0.86 ± 0.4 V and 4.25 ± 1 mV, respectively, at 18 months of follow‐up.</p> </sec> <sec id="jce12579-sec-0030"<abstract abstract-type="main"> <title>Preclinical Feasibility of a Leadless Pacemaker</title> <sec id="jce12579-sec-0010" sec-type="section"> <title>Introduction</title> <p>In this <italic>in vivo</italic> ovine study, we describe the course of pacing and sensing parameters during follow‐up as well as the gross and histopathological features at the implant site of the recently described leadless cardiac pacemaker (LCP).</p> </sec> <sec id="jce12579-sec-0020" sec-type="section"> <title>Methods and Results</title> <p>All sheep underwent LCP implantation in the right ventricular (RV) apex. Serial pacing/sensing thresholds were assessed. In the first cohort, 11 animals were followed up for a period of 3 months, followed by necropsy. In the second cohort, 7 additional sheep were followed for a period of up to 18 months. Mean pacing and sensing thresholds in the 3‐month cohort were 1.0 ± 0.5 V and 9.0 ± 3.7 mV at implant, and 0.7 ± 0.2 V and 8.1 ± 3.9 mV at 90 days, respectively. At necropsy at 3 months, all devices were securely implanted at the RV apex without extrusion of helix beyond the RV wall. Besides endocardial reactive changes in the RV apex surrounding the distal portion of the LCP, there were no other grossly visible abnormalities. In the second cohort (7 sheep), mean pacing and sensing thresholds were 1.0 ± 0.5 V and 9.9 ± 3.8 mV at implant, and 0.86 ± 0.4 V and 4.25 ± 1 mV, respectively, at 18 months of follow‐up.</p> </sec> <sec id="jce12579-sec-0030" sec-type="section"> <title>Conclusion</title> <p>We demonstrate that after implantation of the LCP, pacing/sensing parameters remain adequate up to 18 months in follow‐up. In addition, pathological changes at the implant site and within the RV are limited in severity at 90 days, supporting the efficacy and safety of this novel approach to pacing.</p> </sec> </abstract> … (more)
- Is Part Of:
- Journal of cardiovascular electrophysiology. Volume 26:Number 3(2015:Mar.)
- Journal:
- Journal of cardiovascular electrophysiology
- Issue:
- Volume 26:Number 3(2015:Mar.)
- Issue Display:
- Volume 26, Issue 3 (2015)
- Year:
- 2015
- Volume:
- 26
- Issue:
- 3
- Issue Sort Value:
- 2015-0026-0003-0000
- Page Start:
- 322
- Page End:
- 328
- Publication Date:
- 2014-12-05
- Subjects:
- Blood vessels -- Physiology -- Periodicals
Electrophysiology -- Periodicals
Heart -- Physiology -- Periodicals
612.1 - Journal URLs:
- http://onlinelibrary.wiley.com/ ↗
- DOI:
- 10.1111/jce.12579 ↗
- Languages:
- English
- ISSNs:
- 1045-3873
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4954.866000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 2981.xml