Evaluation of a power injection system in the 7‐Tesla MRI environment. Issue 1 (March 2015)
- Record Type:
- Journal Article
- Title:
- Evaluation of a power injection system in the 7‐Tesla MRI environment. Issue 1 (March 2015)
- Main Title:
- Evaluation of a power injection system in the 7‐Tesla MRI environment
- Authors:
- Dula, Adrienne N.
Shellock, Frank G. - Abstract:
- <abstract abstract-type="main"> <title>ABSTRACT</title> <p>The injection of contrast media enables the visualization of various pathologies through X‐ray, computed tomography, magnetic resonance imaging (MRI), or other medical imaging modalities. Safety and functionality of power injector systems must be evaluated at high magnetic fields (7‐Tesla). The purpose of this study was to evaluate MRI issues for a power injection system (OptiStar Elite MR Contrast Delivery System, Mallinckrodt Pharmaceuticals, Hazelwood, MO) at 7‐Tesla including magnetic field interaction and 31 injector performance tests. Additionally, effects of operation of the power injection system on MR images were evaluated through observation of image artifacts and signal to noise analysis. The components of the power injection system (powerhead, syringes) exhibited no translational attraction when close to 7‐Tesla MR system. Translational attraction was detected for the power supply, power control, and cables, therefore there is a requirement to maintain these components fixed outside of the 0.1000 Tesla line. 7‐Tesla MR scanning did not impact power injection performance. The power injection system produced no discernible image artifacts or degradation when the power controls and powerhead were positioned in the worst‐case position. The power injection system tested for MRI issues passed evaluation for safety and operation at 7‐Tesla and is labeled MR Conditional for use at 7‐T. © 2015 Wiley Periodicals,<abstract abstract-type="main"> <title>ABSTRACT</title> <p>The injection of contrast media enables the visualization of various pathologies through X‐ray, computed tomography, magnetic resonance imaging (MRI), or other medical imaging modalities. Safety and functionality of power injector systems must be evaluated at high magnetic fields (7‐Tesla). The purpose of this study was to evaluate MRI issues for a power injection system (OptiStar Elite MR Contrast Delivery System, Mallinckrodt Pharmaceuticals, Hazelwood, MO) at 7‐Tesla including magnetic field interaction and 31 injector performance tests. Additionally, effects of operation of the power injection system on MR images were evaluated through observation of image artifacts and signal to noise analysis. The components of the power injection system (powerhead, syringes) exhibited no translational attraction when close to 7‐Tesla MR system. Translational attraction was detected for the power supply, power control, and cables, therefore there is a requirement to maintain these components fixed outside of the 0.1000 Tesla line. 7‐Tesla MR scanning did not impact power injection performance. The power injection system produced no discernible image artifacts or degradation when the power controls and powerhead were positioned in the worst‐case position. The power injection system tested for MRI issues passed evaluation for safety and operation at 7‐Tesla and is labeled MR Conditional for use at 7‐T. © 2015 Wiley Periodicals, Inc. Int J Imaging Syst Technol, 25, 50–55, 2015</p> </abstract> … (more)
- Is Part Of:
- International journal of imaging systems and technology. Volume 25:Issue 1(2015:Mar.)
- Journal:
- International journal of imaging systems and technology
- Issue:
- Volume 25:Issue 1(2015:Mar.)
- Issue Display:
- Volume 25, Issue 1 (2015)
- Year:
- 2015
- Volume:
- 25
- Issue:
- 1
- Issue Sort Value:
- 2015-0025-0001-0000
- Page Start:
- 50
- Page End:
- 55
- Publication Date:
- 2015-03
- Subjects:
- Imaging systems -- Periodicals
Image processing -- Periodicals
621.367 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1098-1098 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/ima.22120 ↗
- Languages:
- English
- ISSNs:
- 0899-9457
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4542.299000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 3725.xml