A Comparative Reliability and Performance Study of Different Stent Designs in Terms of Mechanical Properties: Foreshortening, Recoil, Radial Force, and Flexibility. Issue 4 (5th March 2013)
- Record Type:
- Journal Article
- Title:
- A Comparative Reliability and Performance Study of Different Stent Designs in Terms of Mechanical Properties: Foreshortening, Recoil, Radial Force, and Flexibility. Issue 4 (5th March 2013)
- Main Title:
- A Comparative Reliability and Performance Study of Different Stent Designs in Terms of Mechanical Properties: Foreshortening, Recoil, Radial Force, and Flexibility
- Authors:
- Kim, Dong Bin
Choi, Hyuk
Joo, Sang Min
Kim, Han Ki
Shin, Jae Hee
Hwang, Min Ho
Choi, Jaesoon
Kim, Dong‐Gon
Lee, Kwang Ho
Lim, Chun Hak
Yoo, Sun Kook
Lee, Heung‐Man
Sun, Kyung - Abstract:
- <abstract abstract-type="main"> <title>Abstract</title> <p>This study seeks to improve the mechanical performance of stents by conducting reliability performance testing and finite element method (FEM)‐based simulations for coronary stents. Three commercially available stent designs and our own new design were tested to measure the factors affecting performance, specifically foreshortening, recoil, radial force, and flexibility. The stents used in the present experiments were 3 mm in working diameter and 18 mm of working length. The results of the experiments indicate that the foreshortening of stents A, B, C, and our new design, D, was equivalent to 2.25, 0.67, 0.46, and 0.41%, respectively. The recoil of stents A, B, C, and D was 6.00, 4.35, 3.50, and 4.36%, respectively. Parallel plate radial force measurements were A, 3.72 ± 0.28 N; B, 3.81 ± 0.32 N; C, 4.35 ± 0.18 N; and D, 4.02 ± 0.24 N. Radial forces determined by applying uniform pressure in the circumferential direction were A, 28.749 ± 0.81 N; B, 32.231 ± 1.80 N; C, 34.522 ± 3.06 N; and D, 42.183 ± 2.84 N. The maximum force of crimped stent at 2.2‐mm deflection was 1.01 ± 0.08 N, 0.82 ± 0.08 N, 0.92 ± 0.12 N, and 0.68 ± 0.07 N for each of stents A, B, C and D. The results of this study enabled us to identify several factors to enhance the performance of stents. In comparing these stents, we found that our design, stent D, which was designed by a collaborative team from seven universities, performed better than the<abstract abstract-type="main"> <title>Abstract</title> <p>This study seeks to improve the mechanical performance of stents by conducting reliability performance testing and finite element method (FEM)‐based simulations for coronary stents. Three commercially available stent designs and our own new design were tested to measure the factors affecting performance, specifically foreshortening, recoil, radial force, and flexibility. The stents used in the present experiments were 3 mm in working diameter and 18 mm of working length. The results of the experiments indicate that the foreshortening of stents A, B, C, and our new design, D, was equivalent to 2.25, 0.67, 0.46, and 0.41%, respectively. The recoil of stents A, B, C, and D was 6.00, 4.35, 3.50, and 4.36%, respectively. Parallel plate radial force measurements were A, 3.72 ± 0.28 N; B, 3.81 ± 0.32 N; C, 4.35 ± 0.18 N; and D, 4.02 ± 0.24 N. Radial forces determined by applying uniform pressure in the circumferential direction were A, 28.749 ± 0.81 N; B, 32.231 ± 1.80 N; C, 34.522 ± 3.06 N; and D, 42.183 ± 2.84 N. The maximum force of crimped stent at 2.2‐mm deflection was 1.01 ± 0.08 N, 0.82 ± 0.08 N, 0.92 ± 0.12 N, and 0.68 ± 0.07 N for each of stents A, B, C and D. The results of this study enabled us to identify several factors to enhance the performance of stents. In comparing these stents, we found that our design, stent D, which was designed by a collaborative team from seven universities, performed better than the commercial stents across all parameter of foreshortening, recoil, radial force, and flexibility.</p> </abstract> … (more)
- Is Part Of:
- Artificial organs. Volume 37:Issue 4(2013:Apr.)
- Journal:
- Artificial organs
- Issue:
- Volume 37:Issue 4(2013:Apr.)
- Issue Display:
- Volume 37, Issue 4 (2013)
- Year:
- 2013
- Volume:
- 37
- Issue:
- 4
- Issue Sort Value:
- 2013-0037-0004-0000
- Page Start:
- 368
- Page End:
- 379
- Publication Date:
- 2013-03-05
- Subjects:
- Artificial organs -- Periodicals
617.956 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1111/(ISSN)1525-1594 ↗
http://www.blackwell-synergy.com/member/institutions/issuelist.asp?journal=aor ↗
http://onlinelibrary.wiley.com/ ↗
http://firstsearch.oclc.org ↗ - DOI:
- 10.1111/aor.12001 ↗
- Languages:
- English
- ISSNs:
- 0160-564X
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 1735.052000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 3993.xml