How precise is the PRECICE compared to the ISKD in intramedullary limb lengthening?. (June 2014)
- Record Type:
- Journal Article
- Title:
- How precise is the PRECICE compared to the ISKD in intramedullary limb lengthening?. (June 2014)
- Main Title:
- How precise is the PRECICE compared to the ISKD in intramedullary limb lengthening?
- Authors:
- Schiedel, Frank M
Vogt, Björn
Tretow, Henning L
Schuhknecht, Britta
Gosheger, Georg
Horter, Melanie J
Rödl, Robert - Abstract:
- <abstract> <title> <x xml:space="preserve">Abstract</x> </title> <p> <bold>Background and purpose</bold> — The PRECICE intramedullary limb lengthening system uses a new technique with a magnetic rod and a motorized external remote controller (ERC) with rotational magnetic field. We evaluated the reliability and safety of the PRECICE system.</p> <p> <bold>Methods</bold> — We compared our preliminary results with PRECICE in 24 patients (26 nails) with the known difficulties in the use of mechanical lengthening devices such as the ISKD. We used the Paley classification for evaluation of problems, obstacles, and complications.</p> <p> <bold>Results</bold> — 2 nails were primarily without function, and 24/26 nails lengthened over the desired distance. Lengthening desired was 38 mm and lengthening obtained was 37 mm. There were 2 nail breakages, 1 in the welding seam and 1 because of a fall that occurred during consolidation. ERC usage was problematic mostly in patients with femoral lengthening. Adjustment of the ERC was necessary in 10 of 24 cases. 15 cases had implant-associated problems, obstacles were seen in 5 cases, and complications were seen in each of 4 cases.</p> <p> <bold>Interpretaion</bold> — The reliability of the PRECICE system is comparable to that of other intramedullary lengthening devices such as the ISKD. The motorized external remote controller and its application by the patients is a weak point of the system and needs strict supervision.</p> </abstract>
- Is Part Of:
- Acta orthopaedica. Volume 85:Number 3(2014)
- Journal:
- Acta orthopaedica
- Issue:
- Volume 85:Number 3(2014)
- Issue Display:
- Volume 85, Issue 3 (2014)
- Year:
- 2014
- Volume:
- 85
- Issue:
- 3
- Issue Sort Value:
- 2014-0085-0003-0000
- Page Start:
- 293
- Page End:
- 298
- Publication Date:
- 2014-06
- Subjects:
- Orthopedics -- Periodicals
616.7005 - Journal URLs:
- http://informahealthcare.com/loi/ort ↗
http://www.tandfonline.com/toc/iort20/current ↗
https://actaorthop.org/actao/index ↗
http://www.tandfonline.com/ ↗
http://journalsonline.tandf.co.uk/app/home/journal.asp?wasp=65168817ff044fea9c5b577f1cfe2186&referrer=parent&backto=linkingpublicationresults, 1:113260, 1 ↗ - DOI:
- 10.3109/17453674.2014.913955 ↗
- Languages:
- English
- ISSNs:
- 1745-3674
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 0642.055000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 3625.xml