Comparable flow cytometry data can be obtained with two types of instruments, Canto II, and Navios. A GEIL study. Issue 12 (21st October 2013)
- Record Type:
- Journal Article
- Title:
- Comparable flow cytometry data can be obtained with two types of instruments, Canto II, and Navios. A GEIL study. Issue 12 (21st October 2013)
- Main Title:
- Comparable flow cytometry data can be obtained with two types of instruments, Canto II, and Navios. A GEIL study
- Authors:
- Solly, Françoise
Rigollet, Lauren
Baseggio, Lucille
Guy, Julien
Borgeot, Jessica
Guérin, Estelle
Debliquis, Agathe
Drenou, Bernard
Campos, Lydia
Lacombe, Francis
Béné, Marie C. - Abstract:
- <abstract abstract-type="main"> <title>Abstract</title> <p>Flow cytometry (FC) instruments settings classically rely on local establishment of photomultipliers (PMT) voltages adapted to the measurements expected to be performed. In the era of multiparameter FC (MFC), it appears more and more desirable that comparable patterns of fluorescence are obtained in different settings. This relies on a harmonization of settings between instruments. Although this has been shown to be feasible within a given brand of flow cytometers, little information is available about broader comparisons in a given center or in a multicenter fashion. Here, we report a two‐phase series of experiments first performed between a Canto II (BD Biosciences) and a Navios (Beckman Coulter) instruments in the same center. PMT values adjusted on the reference instrument (RI) Canto II were used to establish target values for PMT settings on the paired Navios practice instrument (PI). This allowed to show the good correlation of all but peaks 1 and 2 of Rainbow<sup>®</sup> beads between RI and PI. Using 4‐ or 8‐color stained leukocytes, the similitude of the settings was further confirmed. A complex set of matrices was then established between five centers all equipped with both instruments. Using Bland &amp; Altman difference comparisons for median fluorescence values, it was shown that using either Rainbow beads or CD16 stained polymorphonuclears to set‐up target values on the RI CantoII, highly superimposable<abstract abstract-type="main"> <title>Abstract</title> <p>Flow cytometry (FC) instruments settings classically rely on local establishment of photomultipliers (PMT) voltages adapted to the measurements expected to be performed. In the era of multiparameter FC (MFC), it appears more and more desirable that comparable patterns of fluorescence are obtained in different settings. This relies on a harmonization of settings between instruments. Although this has been shown to be feasible within a given brand of flow cytometers, little information is available about broader comparisons in a given center or in a multicenter fashion. Here, we report a two‐phase series of experiments first performed between a Canto II (BD Biosciences) and a Navios (Beckman Coulter) instruments in the same center. PMT values adjusted on the reference instrument (RI) Canto II were used to establish target values for PMT settings on the paired Navios practice instrument (PI). This allowed to show the good correlation of all but peaks 1 and 2 of Rainbow<sup>®</sup> beads between RI and PI. Using 4‐ or 8‐color stained leukocytes, the similitude of the settings was further confirmed. A complex set of matrices was then established between five centers all equipped with both instruments. Using Bland &amp; Altman difference comparisons for median fluorescence values, it was shown that using either Rainbow beads or CD16 stained polymorphonuclears to set‐up target values on the RI CantoII, highly superimposable results could be obtained on all 9 PI. The latter were obtained using Rainbow beads or Compbeads<sup>®</sup> for comparisons. In summary, this two‐phase study demonstrates the feasibility of different methods allowing for a robust harmonization of settings for MFC. © 2013 International Society for Advancement of Cytometry</p> </abstract> … (more)
- Is Part Of:
- Cytometry. Volume 83:Issue 12(2013:Dec.)
- Journal:
- Cytometry
- Issue:
- Volume 83:Issue 12(2013:Dec.)
- Issue Display:
- Volume 83, Issue 12 (2013)
- Year:
- 2013
- Volume:
- 83
- Issue:
- 12
- Issue Sort Value:
- 2013-0083-0012-0000
- Page Start:
- 1066
- Page End:
- 1072
- Publication Date:
- 2013-10-21
- Subjects:
- Flow cytometry -- Periodicals
Imaging systems in biology -- Periodicals
Imaging systems in medicine -- Periodicals
Diagnostic imaging -- Periodicals
571.605 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1552-4930 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/cyto.a.22404 ↗
- Languages:
- English
- ISSNs:
- 1552-4922
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3506.855100
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 3158.xml