Histological and biomechanical analysis of porous additive manufactured implants made by direct metal laser sintering: A pilot study in sheep. Issue 7 (6th April 2013)
- Record Type:
- Journal Article
- Title:
- Histological and biomechanical analysis of porous additive manufactured implants made by direct metal laser sintering: A pilot study in sheep. Issue 7 (6th April 2013)
- Main Title:
- Histological and biomechanical analysis of porous additive manufactured implants made by direct metal laser sintering: A pilot study in sheep
- Authors:
- Stübinger, Stefan
Mosch, Isabel
Robotti, Pierfrancesco
Sidler, Michéle
Klein, Karina
Ferguson, Stephen J.
von, Brigitte - Abstract:
- <abstract abstract-type="main"> <title> <x xml:space="preserve">Abstract</x> </title> <p>Objectives: It was the aim of this study to analyze osseointegrative properties of porous additive manufactured titanium implants made by direct metal laser sintering in a sheep model after an implantation period of 2 and 8 weeks.</p> <p>Material and Methods: Three different types of implants were placed in the pelvis of six sheep. In each sheep were placed three standard machined (M), three sandblasted and etched (SE), and three porous additive manufactured (AM) implants. Of these three implants (one per type) were examined histologically and six implants were tested biomechanically. Additionally a semiquantitative histomorphometrical and qualitative fluorescent microscopic analysis were performed.</p> <p>Results: After 2 and 8 weeks bone‐to‐implant‐contact (BIC) values of the AM surface (2w: 20.49% ± 5.18%; 8w: 43.91% ± 9.69%) revealed no statistical significant differences in comparison to the M (2w: 20.33% ± 11.50%; 8w: 25.33% ± 4.61%) and SE (2w: 43.67 ± 12.22%; 8w: 53.33 ± 8.96%) surfaces. AM surface showed the highest increase of the BIC between the two observation time points. Considering the same implantation period histomorphometry and fluorescent labelling disclosed no significant differences in the bone surrounding the three implants groups. In contrast Removal‐torque‐test showed a significant improve in fixation strength (<italic>P</italic> ≤ 0.001) for the AM (1891.82 ±<abstract abstract-type="main"> <title> <x xml:space="preserve">Abstract</x> </title> <p>Objectives: It was the aim of this study to analyze osseointegrative properties of porous additive manufactured titanium implants made by direct metal laser sintering in a sheep model after an implantation period of 2 and 8 weeks.</p> <p>Material and Methods: Three different types of implants were placed in the pelvis of six sheep. In each sheep were placed three standard machined (M), three sandblasted and etched (SE), and three porous additive manufactured (AM) implants. Of these three implants (one per type) were examined histologically and six implants were tested biomechanically. Additionally a semiquantitative histomorphometrical and qualitative fluorescent microscopic analysis were performed.</p> <p>Results: After 2 and 8 weeks bone‐to‐implant‐contact (BIC) values of the AM surface (2w: 20.49% ± 5.18%; 8w: 43.91% ± 9.69%) revealed no statistical significant differences in comparison to the M (2w: 20.33% ± 11.50%; 8w: 25.33% ± 4.61%) and SE (2w: 43.67 ± 12.22%; 8w: 53.33 ± 8.96%) surfaces. AM surface showed the highest increase of the BIC between the two observation time points. Considering the same implantation period histomorphometry and fluorescent labelling disclosed no significant differences in the bone surrounding the three implants groups. In contrast Removal‐torque‐test showed a significant improve in fixation strength (<italic>P</italic> ≤ 0.001) for the AM (1891.82 ± 308, 44 Nmm) surface after eight weeks in comparison to the M (198.93±88, 04 Nmm) and SE (730.08 ± 151, 89 Nmm) surfaces.</p> <p>Conclusion: All three surfaces (M, SE, and AM) showed sound osseointegration. AM implants may offer a possible treatment option in clinics for patients with compromised bone situations. © 2013 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater 101B: 1154–1163, 2013.</p> </abstract> … (more)
- Is Part Of:
- Journal of biomedical materials research. Volume 101:Issue 7(2013:Oct.)
- Journal:
- Journal of biomedical materials research
- Issue:
- Volume 101:Issue 7(2013:Oct.)
- Issue Display:
- Volume 101, Issue 7 (2013)
- Year:
- 2013
- Volume:
- 101
- Issue:
- 7
- Issue Sort Value:
- 2013-0101-0007-0000
- Page Start:
- 1154
- Page End:
- 1163
- Publication Date:
- 2013-04-06
- Subjects:
- Biomedical materials -- Periodicals
610.28 - Journal URLs:
- http://onlinelibrary.wiley.com/ ↗
- DOI:
- 10.1002/jbm.b.32925 ↗
- Languages:
- English
- ISSNs:
- 1552-4973
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4953.725000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 4235.xml