- <abstract abstract-type="main"> <title> <x xml:space="preserve">Abstract</x> </title> <sec id="jmri23923-sec-0001" sec-type="section"> <title>Purpose</title> <p>To compare the repeatability of γ‐aminobutyric acid (GABA) measurements using <italic>J</italic>‐difference editing, before and after spectral realignment—a technique which has previously been demonstrated to improve the quality of <italic>J</italic>‐difference GABA spectra.</p> </sec> <sec id="jmri23923-sec-0002" sec-type="section"> <title>Materials and Methods</title> <p>We performed in vivo measurements in three brain regions (occipital, sensorimotor, and dorsolateral prefrontal cortex [DLPFC]), and analyzed these using alternative alignment approaches to evaluate the impact of alignment on repeatability: "Independent alignment" (aligning each subspectrum independently) and "Pairwise alignment" (aligning each <italic>on</italic> and <italic>off</italic> subspectrum as a pair) were compared.</p> </sec> <sec id="jmri23923-sec-0003" sec-type="section"> <title>Results</title> <p>Pairwise alignment improved the group mean coefficient of variation in all regions; 0.4% in occipital, 1.1% in sensorimotor, and 1.1% in DLPFC. Independent alignment resulted in subtraction artifacts in the majority of cases, and increased the coefficient of variation in the DLPFC by 9.4%. Simulations demonstrate that the GABA quantification error in datasets with high <italic>B</italic><sub><italic>0</italic></sub> drift, is 4.5% without<abstract abstract-type="main"> <title> <x xml:space="preserve">Abstract</x> </title> <sec id="jmri23923-sec-0001" sec-type="section"> <title>Purpose</title> <p>To compare the repeatability of γ‐aminobutyric acid (GABA) measurements using <italic>J</italic>‐difference editing, before and after spectral realignment—a technique which has previously been demonstrated to improve the quality of <italic>J</italic>‐difference GABA spectra.</p> </sec> <sec id="jmri23923-sec-0002" sec-type="section"> <title>Materials and Methods</title> <p>We performed in vivo measurements in three brain regions (occipital, sensorimotor, and dorsolateral prefrontal cortex [DLPFC]), and analyzed these using alternative alignment approaches to evaluate the impact of alignment on repeatability: "Independent alignment" (aligning each subspectrum independently) and "Pairwise alignment" (aligning each <italic>on</italic> and <italic>off</italic> subspectrum as a pair) were compared.</p> </sec> <sec id="jmri23923-sec-0003" sec-type="section"> <title>Results</title> <p>Pairwise alignment improved the group mean coefficient of variation in all regions; 0.4% in occipital, 1.1% in sensorimotor, and 1.1% in DLPFC. Independent alignment resulted in subtraction artifacts in the majority of cases, and increased the coefficient of variation in the DLPFC by 9.4%. Simulations demonstrate that the GABA quantification error in datasets with high <italic>B</italic><sub><italic>0</italic></sub> drift, is 4.5% without alignment, but &lt;1% with optimal alignment.</p> </sec> <sec id="jmri23923-sec-0004" sec-type="section"> <title>Conclusion</title> <p>Pairwise alignment improves the repeatability of GABA spectroscopy data. However, independently aligning all <italic>on</italic> and <italic>off</italic> subspectra can lead to artifacts and worse repeatability when compared with nonaligned data. J. Magn. Reson. Imaging 2013;38:970–975. © 2012 Wiley Periodicals, Inc.</p> </sec> </abstract> … (more)
- Is Part Of:
- Journal of magnetic resonance imaging. Volume 38:Issue 4(2013)
- Journal:
- Journal of magnetic resonance imaging
- Issue:
- Volume 38:Issue 4(2013)
- Issue Display:
- Volume 38, Issue 4 (2013)
- Year:
- 2013
- Volume:
- 38
- Issue:
- 4
- Issue Sort Value:
- 2013-0038-0004-0000
- Page Start:
- 970
- Page End:
- 975
- Publication Date:
- 2012-11-27
- Subjects:
- Magnetic resonance imaging -- Periodicals
616 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1522-2586 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/jmri.23923 ↗
- Languages:
- English
- ISSNs:
- 1053-1807
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5010.791000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 3757.xml