Comparison of methods for determining the partition coefficient of gadolinium in the myocardium using T1 mapping. Issue 1 (29th November 2012)
- Record Type:
- Journal Article
- Title:
- Comparison of methods for determining the partition coefficient of gadolinium in the myocardium using T1 mapping. Issue 1 (29th November 2012)
- Main Title:
- Comparison of methods for determining the partition coefficient of gadolinium in the myocardium using T1 mapping
- Authors:
- Salerno, Michael
Janardhanan, Rajesh
Jiji, Ronny S.
Brooks, Jeremy
Adenaw, Nebiyu
Mehta, Bhairav
Yang, Yang
Antkowiak, Patrick
Kramer, Christopher M.
Epstein, Frederick H. - Abstract:
- <abstract abstract-type="main"> <title> <x xml:space="preserve">Abstract</x> </title> <sec id="jmri23875-sec-0001" sec-type="section"> <title>Purpose</title> <p>To develop and validate modified Look‐Locker (MOLLI) protocols to generate myocardial T<sub>1</sub> maps within clinically acceptable breath‐hold durations and to compare partition coefficients (λ) of gadolinium (Gd)‐DTPA determined from either bolus injection (BI) or continuous infusion (CI) techniques.</p> </sec> <sec id="jmri23875-sec-0002" sec-type="section"> <title>Materials and Methods</title> <p>T<sub>1</sub> mapping was performed in phantoms and in 10 volunteers on a 1.5T scanner using the standard (3‐3‐5) MOLLI technique and two MOLLI schemes with shorter breath‐hold durations. Imaging was performed precontrast and every 5 minutes following a bolus of 0.1 mmol/kg Gd‐DTPA and a 15‐minute delayed continuous infusion of 0.001 mmol/kg Gd‐DTPA until equilibrium T<sub>1</sub> in the myocardium was achieved to enable direct comparison of T<sub>1</sub> relaxation times between techniques and λ's between the BI and CI methods.</p> </sec> <sec id="jmri23875-sec-0003" sec-type="section"> <title>Results</title> <p>There was good agreement of T<sub>1</sub> values between the 3‐3‐5 standard MOLLI protocol and the modified 3‐5 MOLLI protocol in both phantom studies over a range of heart rates and in human subjects. Both MOLLI protocols produced similar measurements of λ using both the BI and CI methods.</p> </sec> <sec<abstract abstract-type="main"> <title> <x xml:space="preserve">Abstract</x> </title> <sec id="jmri23875-sec-0001" sec-type="section"> <title>Purpose</title> <p>To develop and validate modified Look‐Locker (MOLLI) protocols to generate myocardial T<sub>1</sub> maps within clinically acceptable breath‐hold durations and to compare partition coefficients (λ) of gadolinium (Gd)‐DTPA determined from either bolus injection (BI) or continuous infusion (CI) techniques.</p> </sec> <sec id="jmri23875-sec-0002" sec-type="section"> <title>Materials and Methods</title> <p>T<sub>1</sub> mapping was performed in phantoms and in 10 volunteers on a 1.5T scanner using the standard (3‐3‐5) MOLLI technique and two MOLLI schemes with shorter breath‐hold durations. Imaging was performed precontrast and every 5 minutes following a bolus of 0.1 mmol/kg Gd‐DTPA and a 15‐minute delayed continuous infusion of 0.001 mmol/kg Gd‐DTPA until equilibrium T<sub>1</sub> in the myocardium was achieved to enable direct comparison of T<sub>1</sub> relaxation times between techniques and λ's between the BI and CI methods.</p> </sec> <sec id="jmri23875-sec-0003" sec-type="section"> <title>Results</title> <p>There was good agreement of T<sub>1</sub> values between the 3‐3‐5 standard MOLLI protocol and the modified 3‐5 MOLLI protocol in both phantom studies over a range of heart rates and in human subjects. Both MOLLI protocols produced similar measurements of λ using both the BI and CI methods.</p> </sec> <sec id="jmri23875-sec-0004" sec-type="section"> <title>Conclusion</title> <p>A reduced breath‐hold MOLLI T<sub>1</sub> mapping protocol combined with the BI method can accurately characterize T<sub>1</sub> and λ in clinically applicable breath‐hold durations without requiring a long equilibrium phase infusion. J. Magn. Reson. Imaging 2013;38:217–224. © 2012 Wiley Periodicals, Inc.</p> </sec> </abstract> … (more)
- Is Part Of:
- Journal of magnetic resonance imaging. Volume 38:Issue 1(2013)
- Journal:
- Journal of magnetic resonance imaging
- Issue:
- Volume 38:Issue 1(2013)
- Issue Display:
- Volume 38, Issue 1 (2013)
- Year:
- 2013
- Volume:
- 38
- Issue:
- 1
- Issue Sort Value:
- 2013-0038-0001-0000
- Page Start:
- 217
- Page End:
- 224
- Publication Date:
- 2012-11-29
- Subjects:
- Magnetic resonance imaging -- Periodicals
616 - Journal URLs:
- http://onlinelibrary.wiley.com/journal/10.1002/(ISSN)1522-2586 ↗
http://onlinelibrary.wiley.com/ ↗ - DOI:
- 10.1002/jmri.23875 ↗
- Languages:
- English
- ISSNs:
- 1053-1807
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 5010.791000
British Library DSC - BLDSS-3PM
British Library HMNTS - ELD Digital store - Ingest File:
- 3755.xml