Clinical validation of the AMTAS automated audiometer. (May 2013)
- Record Type:
- Journal Article
- Title:
- Clinical validation of the AMTAS automated audiometer. (May 2013)
- Main Title:
- Clinical validation of the AMTAS automated audiometer
- Authors:
- Eikelboom, Robert H.
Swanepoel, De Wet
Motakef, Shahpar
Upson, Gemma S. - Abstract:
- <abstract> <title>Abstract</title> <p> <italic>Objective:</italic> To validate the air- and bone-conduction AMTAS automated audiometry system. <italic>Design:</italic> Prospective study. Test-retest reliability was determined by assessing adults with AMTAS air- and bone-conduction audiometry. Accuracy was determined by comparing AMTAS and manual audiometry conducted on adults. AMTAS testing was conducted in a quiet room and manual audiometry in a sound booth. <italic>Study sample:</italic> Ten participants for test-retest reliability tests and 44 participants to determine accuracy were included. Participants had varying degrees of hearing loss. <italic>Results:</italic> For test-retest reliability the overall difference in air-conduction hearing thresholds (n = 119) was 0.5 dB. The spread of differences (standard deviation of absolute differences) was 4.9 dB. For bone-conduction thresholds (n = 99) the overall difference was − 0.2 dB, and the spread of differences 4.5 dB. For accuracy the overall difference in air-conduction hearing thresholds (n = 509) between the two techniques was 0.1 dB. The spread of differences was 6.4 dB. For bone-conduction thresholds (n = 295) the overall difference was 0 dB, and the spread of differences 7.7 dB. <italic>Conclusions:</italic> Variations between air- and bone-conduction audiometry for automated and manual audiometry were within normally accepted limits for audiometry. However, AMTAS thresholds were elevated but not significantly<abstract> <title>Abstract</title> <p> <italic>Objective:</italic> To validate the air- and bone-conduction AMTAS automated audiometry system. <italic>Design:</italic> Prospective study. Test-retest reliability was determined by assessing adults with AMTAS air- and bone-conduction audiometry. Accuracy was determined by comparing AMTAS and manual audiometry conducted on adults. AMTAS testing was conducted in a quiet room and manual audiometry in a sound booth. <italic>Study sample:</italic> Ten participants for test-retest reliability tests and 44 participants to determine accuracy were included. Participants had varying degrees of hearing loss. <italic>Results:</italic> For test-retest reliability the overall difference in air-conduction hearing thresholds (n = 119) was 0.5 dB. The spread of differences (standard deviation of absolute differences) was 4.9 dB. For bone-conduction thresholds (n = 99) the overall difference was − 0.2 dB, and the spread of differences 4.5 dB. For accuracy the overall difference in air-conduction hearing thresholds (n = 509) between the two techniques was 0.1 dB. The spread of differences was 6.4 dB. For bone-conduction thresholds (n = 295) the overall difference was 0 dB, and the spread of differences 7.7 dB. <italic>Conclusions:</italic> Variations between air- and bone-conduction audiometry for automated and manual audiometry were within normally accepted limits for audiometry. However, AMTAS thresholds were elevated but not significantly different compared to other contemporary studies that included an automated audiometer.</p> </abstract> … (more)
- Is Part Of:
- International journal of audiology. Volume 52:Number 5(2013:May)
- Journal:
- International journal of audiology
- Issue:
- Volume 52:Number 5(2013:May)
- Issue Display:
- Volume 52, Issue 5 (2013)
- Year:
- 2013
- Volume:
- 52
- Issue:
- 5
- Issue Sort Value:
- 2013-0052-0005-0000
- Page Start:
- 342
- Page End:
- 349
- Publication Date:
- 2013-05
- Subjects:
- Audiology -- Periodicals
Hearing disorders -- Periodicals
Deafness -- Prevention -- Periodicals
Hearing Disorders -- Periodicals
Hearing -- Periodicals
617.8 - Journal URLs:
- http://informahealthcare.com/loi/ija ↗
http://www.tandfonline.com/loi/iija20?open=54&repitition=0 ↗
http://informahealthcare.com ↗
http://journalsonline.tandf.co.uk/app/home/journal.asp?wasp=e54fd9ec35a8443595d2fe2a284d67dd&referrer=parent&backto=linkingpublicationresults, 1:112274, 1 ↗ - DOI:
- 10.3109/14992027.2013.769065 ↗
- Languages:
- English
- ISSNs:
- 1499-2027
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 4542.115000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 4392.xml