Fractured zirconia implants and related implant designs: scanning electron microscopy analysis. Issue 5 (26th January 2012)
- Record Type:
- Journal Article
- Title:
- Fractured zirconia implants and related implant designs: scanning electron microscopy analysis. Issue 5 (26th January 2012)
- Main Title:
- Fractured zirconia implants and related implant designs: scanning electron microscopy analysis
- Authors:
- Osman, Reham B.
Ma, Sunyoung
Duncan, Warwick
De Silva, Rohana K.
Siddiqi, Allauddin
Swain, Michael V. - Abstract:
- <abstract abstract-type="main" id="clr2411-abs-0001"> <title>Abstract</title> <sec id="clr2411-sec-0001" sec-type="section"> <title>Objectives</title> <p>Two fractured one‐piece experimental (commercially unavailable) zirconia implants were analyzed using scanning electron microscope (SEM) analysis to identify failure origins and aid in understanding the failure mechanisms. Modifications to the zirconia implant design are suggested to minimize such fracture incidences.</p> </sec> <sec id="clr2411-sec-0002" sec-type="section"> <title>Materials and methods</title> <p>Two zirconia implants fractured during the final torquing in the maxillary ridge using the prescribed hand torque wrench. The implants were subsequently retrieved and prepared for optical and SEM evaluation. Critical attention was given to the fractography (crack morphology) of the fractured implants to identify the fracture origin.</p> </sec> <sec id="clr2411-sec-0003" sec-type="section"> <title>Results</title> <p>Events related to initiation and propagation of the crack front could be detected from the morphology of the fractured surfaces. Unfavorable torque and bending forces applied on the implant during surgical placement and the inherent flaws in the material may have resulted in crack initiation and implant failure.</p> </sec> <sec id="clr2411-sec-0004" sec-type="section"> <title>Conclusions</title> <p>Caution must be exercised when placing zirconia implants in dense bone sites. Modification of surgical<abstract abstract-type="main" id="clr2411-abs-0001"> <title>Abstract</title> <sec id="clr2411-sec-0001" sec-type="section"> <title>Objectives</title> <p>Two fractured one‐piece experimental (commercially unavailable) zirconia implants were analyzed using scanning electron microscope (SEM) analysis to identify failure origins and aid in understanding the failure mechanisms. Modifications to the zirconia implant design are suggested to minimize such fracture incidences.</p> </sec> <sec id="clr2411-sec-0002" sec-type="section"> <title>Materials and methods</title> <p>Two zirconia implants fractured during the final torquing in the maxillary ridge using the prescribed hand torque wrench. The implants were subsequently retrieved and prepared for optical and SEM evaluation. Critical attention was given to the fractography (crack morphology) of the fractured implants to identify the fracture origin.</p> </sec> <sec id="clr2411-sec-0003" sec-type="section"> <title>Results</title> <p>Events related to initiation and propagation of the crack front could be detected from the morphology of the fractured surfaces. Unfavorable torque and bending forces applied on the implant during surgical placement and the inherent flaws in the material may have resulted in crack initiation and implant failure.</p> </sec> <sec id="clr2411-sec-0004" sec-type="section"> <title>Conclusions</title> <p>Caution must be exercised when placing zirconia implants in dense bone sites. Modification of surgical protocols for the intended implant site may be necessary. Improvement in design features specific to zirconia implants, and strict quality control during manufacture is essential to minimize the likelihood of fracture.</p> </sec> </abstract> … (more)
- Is Part Of:
- Clinical oral implants research. Volume 24:Issue 5(2013:May)
- Journal:
- Clinical oral implants research
- Issue:
- Volume 24:Issue 5(2013:May)
- Issue Display:
- Volume 24, Issue 5 (2013)
- Year:
- 2013
- Volume:
- 24
- Issue:
- 5
- Issue Sort Value:
- 2013-0024-0005-0000
- Page Start:
- 592
- Page End:
- 597
- Publication Date:
- 2012-01-26
- Subjects:
- Dental implants -- Research -- Periodicals
617.69 - Journal URLs:
- http://onlinelibrary.wiley.com/ ↗
- DOI:
- 10.1111/j.1600-0501.2011.02411.x ↗
- Languages:
- English
- ISSNs:
- 0905-7161
- Deposit Type:
- Legaldeposit
- View Content:
- Available online (eLD content is only available in our Reading Rooms) ↗
- Physical Locations:
- British Library DSC - 3286.318000
British Library DSC - BLDSS-3PM
British Library STI - ELD Digital store - Ingest File:
- 3891.xml